TopPette Mechanical Pipette is a half autoclavable pipette, with wide range of volume from 0.1μL to 10mL. Single channel, 8-channel, 12-channel adjustable volume and single channel fixed volume are options.
Main Features
- Light weight, ergonomic, low force design
- Digital display clearly reads volume setting
- The pipettes cover volume range of 0.1μL to 10mL
- Easy to calibrate and maintain with tool supplied
- Design helps avoid repetitive strain injuries
- Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
- The lower section can be autoclaved
- Ultrassay Pipettes can be stopped with an on/off switch and cable. This can be used with coagulation analyzers or any other instruments that require precise timing.
Technical Specification
Single-Channel Adjustable Volume Pipettes
| Volume Range | Increment | Test Volume | Systematic Error | Radomn Error | ||
| μL | μL | μL | μL | % | μL | % |
| 0.1~2.5 | 0.05 | 2.5 | ±2.5 | ±0.0625 | ±2.0 | ±0.05 |
| 1.25 | ±3.0 | ±0.0375 | ±3.0 | ±0.0375 | ||
| 0.25 | ±12.0 | ±0.03 | ±6.0 | ±0.015 | ||
| 0.5~10 | 0.1 | 10 | ±1.0 | ±0.1 | ±0.8 | ±0.08 |
| 5 | ±1.5 | ±0.075 | ±1.5 | ±0.075 | ||
| 1 | ±2.5 | ±0.025 | ±1.5 | ±0.015 | ||
| 2~20 | 0.5 | 20 | ±0.9 | ±0.18 | ±0.04 | ±0.015 |
| 10 | ±1.2 | ±0.12 | ±1.0 | ±0.08 | ||
| 2 | ±3.0 | ±0.06 | ±2.0 | ±0.1 | ||
| 5~50 | 0.5 | 50 | ±0.6 | ±0.3 | ±0.3 | ±0.15 |
| 25 | ±0.9 | ±0.225 | ±0.6 | ±0.15 | ||
| 5 | ±2.0 | ±0.1 | ±2.0 | ±0.1 | ||
| 10~100 | 1 | 100 | ±0.8 | ±0.8 | ±0.15 | ±0.15 |
| 50 | ±1.0 | ±0.5 | ±0.4 | ±0.2 | ||
| 10 | ±3.0 | ±0.3 | ±1.5 | ±0.15 | ||
| 20~200 | 1 | 200 | ±0.6 | ±1.2 | ±0.15 | ±0.3 |
| 100 | ±0.8 | ±0.8 | ±0.3 | ±0.3 | ||
| 20 | ±3.0 | ±0.6 | ±1.0 | ±0.2 | ||
| 50~200 | 1 | 200 | ±0.6 | ±1.2 | ±0.15 | ±0.3 |
| 100 | ±0.8 | ±0.8 | ±0.3 | ±0.3 | ||
| 50 | ±1.0 | ±0.5 | ±0.4 | ±0.2 | ||
| 100~1000 | 5 | 1000 | ±0.6 | ±6 | ±2.0 | ±2 |
| 500 | ±0.7 | ±3.5 | ±0.2 | ±1.25 | ||
| 100 | ±2.0 | ±2 | ±0.25 | ±0.7 | ||
| 200~1000 | 5 | 1000 | ±0.6 | ±6 | ±2.0 | ±2 |
| 500 | ±0.7 | ±3.5 | ±0.2 | ±1.25 | ||
| 200 | ±0.9 | ±1.8 | ±0.3 | ±0.6 | ||
| 500~5000 | 50 | 5000 | ±0.5 | ±25 | ±0.15 | ±7.5 |
| 2500 | ±0.6 | ±15 | ±0.3 | ±7.5 | ||
| 500 | ±0.7 | ±3.5 | ±0.3 | ±1.5 | ||
| 1000~5000 | 50 | 5000 | ±0.5 | ±25 | ±0.15 | ±7.5 |
| 2500 | ±0.6 | ±15 | ±0.3 | ±7.5 | ||
| 1000 | ±0.7 | ±7 | ±0.3 | ±3 | ||
| 1000~10000 | 100 | 10000 | ±0.6 | ±60 | ±0.2 | ±20 |
| 5000 | ±1.2 | ±60 | ±0.3 | ±15 | ||
| 1000 | ±3.0 | ±30 | ±0.6 | ±6 | ||
| 2000~10000 | 100 | 10000 | ±0.6 | ±60 | ±0.2 | ±20 |
| 5000 | ±1.2 | ±60 | ±0.3 | ±15 | ||
| 2000 | ±3.0 | ±60 | ±0.6 | ±12 | ||
8-Channel Adjustable Volume Pipettes
| Volume Range | Increment | Test Volume | Systematic Error | Radomn Error | ||
| μL | μL | μL | μL | % | μL | % |
| 0.5~10 | 0.1 | 10 | ±1.5 | ±0.15 | ±1.5 | ±0.15 |
| 5 | ±2.5 | ±0.125 | ±2.5 | ±0.125 | ||
| 1 | ±4.0 | ±0.04 | ±4.0 | ±0.04 | ||
| 5~50 | 0.5 | 50 | ±1.0 | ±0.5 | ±0.5 | ±0.25 |
| 25 | ±1.5 | ±0.375 | ±1.0 | ±0.25 | ||
| 5 | ±3.0 | ±0.15 | ±2.0 | ±0.1 | ||
| 10~100 | 1 | 100 | ±0.8 | ±0.8 | ±0.3 | ±0.3 |
| 50 | ±1.0 | ±0.5 | ±0.8 | ±0.4 | ||
| 10 | ±3.0 | ±0.3 | ±2.0 | ±0.2 | ||
| 50~300 | 5 | 300 | ±0.7 | ±2.1 | ±0.25 | ±0.75 |
| 150 | ±1.0 | ±1.5 | ±0.5 | ±0.75 | ||
| 30 | ±1.5 | ±0.75 | ±0.8 | ±0.4 | ||
12-Channel Adjustable Volume Pipettes
| Volume Range | Increment | Test Volume | Systematic Error | Radomn Error | ||
| μL | μL | μL | μL | % | μL | % |
| 0.5~10 | 0.1 | 10 | ±1.5 | ±0.15 | ±1.5 | ±0.15 |
| 5 | ±2.5 | ±0.125 | ±2.5 | ±0.125 | ||
| 1 | ±4.0 | ±0.04 | ±4.0 | ±0.04 | ||
| 5~50 | 0.5 | 50 | ±1.0 | ±0.5 | ±0.5 | ±0.25 |
| 25 | ±1.5 | ±0.375 | ±1.0 | ±0.25 | ||
| 5 | ±3.0 | ±0.15 | ±2.0 | ±0.1 | ||
| 10~100 | 1 | 100 | ±0.8 | ±0.8 | ±0.3 | ±0.3 |
| 50 | ±1.0 | ±0.5 | ±0.8 | ±0.4 | ||
| 10 | ±3.0 | ±0.3 | ±2.0 | ±0.2 | ||
| 50~300 | 5 | 300 | ±0.7 | ±2.1 | ±0.25 | ±0.75 |
| 150 | ±1.0 | ±1.5 | ±0.5 | ±0.75 | ||
| 30 | ±1.5 | ±0.75 | ±0.8 | ±0.4 | ||
Single-Channel Fixed Volume Pipettes
| Volume Range | Increment | Test Volume | Systematic Error | Radomn Error | ||
| μL | μL | μL | μL | % | μL | % |
| 5 | – | 5 | ±1.3 | ±0.065 | ±1.2 | ±0.06 |
| 10 | – | 10 | ±0.8 | ±0.08 | ±0.8 | ±0.08 |
| 20 | – | 20 | ±0.6 | ±0.12 | ±0.5 | ±0.1 |
| 25 | – | 25 | ±0.5 | ±0.125 | ±0.3 | ±0.075 |
| 50 | – | 50 | ±0.5 | ±0.25 | ±0.3 | ±0.15 |
| 100 | – | 100 | ±0.5 | ±0.5 | ±0.3 | ±0.3 |
| 200 | – | 200 | ±0.4 | ±0.8 | ±0.2 | ±0.4 |
| 250 | – | 250 | ±0.4 | ±1.0 | ±0.2 | ±0.5 |
| 500 | – | 500 | ±0.3 | ±1.5 | ±0.2 | ±1.0 |
| 1000 | – | 1000 | ±0.3 | ±3.0 | ±0.2 | ±2.0 |
| 2000 | – | 2000 | ±0.3 | ±6.0 | ±0.15 | ±3.0 |
| 5000 | – | 5000 | ±0.3 | ±15 | ±0.15 | ±7.5 |
*Ultrassay BioTech Co., Ltd. reserves the right of final interpretation of different specifications.