eQ9600 Real-time PCR System
Feature:
Ultrassay eQ9600 Real-time PCR system is a 96-well high-throughput real-time fluorescent quantitative PCR detection system released by ultrassay following the eQ1600 series. This product adpots the DDSOM two-dimensional scanning photoelectric module technology and thermal system design, the unique technology of ultrassay achieving product characteristics of strong photolectric signal, high sensitivity, low background noise and light weight. Coupled with powerful and friendly system software, it can be used for multiple gene detection, quantitative analysis, SNP analysis, melting curve analysis and other experiments.
This system is suitable for polymerase chain reaction and quantitative detection in immunology, human genome engineering, forensic medicine, oncology, tissue and population biology, paleontology, zoology, botany and other fields.
Advantages:
1. Touch screen control, with software, efficient and convenient.
2. Simple operation interface, powerful software function.
3. Strong fluorescence signal, low background noise, high sensitivity.
4. Unique design of block, rapid temperature change, good uniformity.
Application scope:
This system is optimized for configuration, especially for sample size, rapid on-site inspection and quarantine, disease prevention and control (such as African swine fever, bird flu, etc.), food safety, agriculture, forestry and biochemistry.
Technical Specification
Sample volume | 96*0.2ml single tube or 8-strip transparent tube (except for white opaque tube), you can mark the tube with markers on the cover of the tube. |
Applicable reagents | This system is an open platform, suitable for various real-time PCR reagents, including fast extraction reagents, non-extraction reagents, or direct amplification reagents and other low-concentration PCR reagent. |
Reaction volume | 5-100ul |
Dynamic range | 1-1010 Copies |
Color wavelength | 320 – 900nm |
Optical sensor | PDM |
Dyes | CH1: FAM, SYBR Green I; CH2: HEX, JOE, VIC, TAMRA; CH3: ROX, CALRED; CH4: CY5, QUASAR CH5: Optional CH6: Optional |
Sensitivity | 1 Copy |
Rn | ≥0.998 |
CV | ≤3% |
Block setting range | 10-99.9ºC |
Temperature control mode | Tube mode, Real temperature of simulated reagent |
Refrigeration | Peltier |
Precision of temperature control | ≤0.1ºC |
Display resolution | 0.1ºC |
Temperature accuracy | ≤±0.2ºC |
Temperature uniformity | ≤0.2ºC |
Max. ramping rate | ≥4ºC/s |
Avg. ramping rate | ≥2.5ºC/s (50ºC – 90ºC) |
Hot lid | 30ºC-105ºC |
Operating system | Windows 7/10 |
Communication Interface | RS232/USB |
Software functions | Built-in two-steps, three-steps, LAMP, RPA methods’ PCR program templates, free extraction reagent PCR program template and other common ones. It has functions of system setting (password permission), experimental parameter settings, sample information input, operation management, data export EXCEL, general view of PCR program, real-time amplification curve display, channel crosstalk correction, criterion setting, automatic report, curve capture, etc. |
Analytical function | Standard curve, relative quantitative curve and melting curve (HRM) |
Operating ambient | Ambient temperature: 10-30ºC, relative humidity: 20-85% RH, |
Input power | 100-240V 50-60Hz 800W |
Dimension | 428x305x202 (A3 paper size) |
Net weight | 12.5kg |